Checks and Adjustments
1705A Spectrum Monitor
5- 25
g.
ADJUST
-- R51, R58, R75, R81, and R82 for a linear display with each
marker on a graticule mark,
±
1 minor Division.
h.
Readjust R35 to place first marker on the extreme left graticule.
NOTE
. Readjustment of R35 may cause a change in linearity.
13. Adjust Readout Range
a.
Turn CENTER FREQUENCY fully counterclockwise.
b. ADJUST
-- R4 (Readout Cal) so the frequency readout displays
900 MHz.
14. Adjust Span/Div Magnification Registration
a.
Set SPAN/DIV to 1 MHz.
b.
Turn the CENTER FREQUENCY control to position any marker under
the readout caret.
c.
Set SPAN/DIV to 10 MHz.
d. ADJUST
-- R27 (Horiz Posn Readout) to position the caret directly over
the marker.
e.
Set SPAN/DIV to 1 MHz. Marker should be under caret
±
1 minor
Division.
f.
Adjust the HORIZONTAL POSITION control to place the caret at the
center of the screen.
g.
If R27 requires adjustment, check step 10 parts a. through e. (R55), and
check step 12 parts a. through d. (R35).
15. Adjust Span/Div Magnifier Range
a.
Set SPAN/DIV to 10 MHz.
b.
Use the HORIZONTAL POSITION control to position the caret to the
center screen graticule line.
c.
Turn the CENTER FREQUENCY control fully clockwise (the readout
should display 2000 MHz
±
50 MHz).
d.
Set the signal generator Frequency to 1800 MHz, and Amplitude to --30
dBm.
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......