Checks and Adjustments
5- 12
1705A Spectrum Monitor
10. Check L--Band Linearity
REQUIREMENT
-- One marker per Division
±
1 minor Division.
a.
Connect the output of the leveled sine wave generator, through the comb
generator, to the 1705A L--BAND INPUT. See Figure 5-4.
SG503
Comb
Generator
VARIABLE
AUTOTRANSFORMER
L--BAND
INPUT
1705A
Figure 5-4: Equipment connections for L- Band checkout
b.
Set the leveled sine wave generator Frequency to 100 MHz, Amplitude
Multiplier to X1, and the Amplitude to 1.5 (8 dBm).
c.
Set the 1705A INPUT to L--BAND.
d.
Set the SPAN/DIV to FULL.
e.
CHECK
-- for 10 frequency markers. (Ignore any sweep start marker.)
f.
CHECK
– that each marker is within one minor Division of a major
graticule line when the 2
nd
marker is on the first graticule line in from
the left. See Figure 5-5.
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......