2. General
28
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
2021-07
2.6.2
Microplate Shaking
The SUNRISE is able to shake the microplate before it is measured. Use external
software (for example: Magellan) to set the shaking modes.
The microplate can also be shaken between each of the kinetic measurement
cycles.
WARNING
WHEN USING A 96 WELL PLATE IN HIGH SHAKE MODE, SPILLAGE
MAY OCCUR IF THE WELLS ARE FILLED WITH MORE THAN 300 µL.
The shaking widths and frequencies for the 4 SUNRISE shake modes are as
follows:
Shake Mode
Shake Width
Shake Frequency
HIGH
NORMAL
LOW
WIDE
2.8 mm
4.4 mm
4.4 mm
14.2 mm
12.3 Hz
9.2 Hz
7.8 Hz
2 Hz
2.7
Instrument Spare Parts and Tools
The table below contains spare parts and tools:
Part Name
•
Halogen Lamp
•
Reader to External Computer Cable
•
QC Pac 1 for SUNRISE
•
QC Pac 2 for SUNRISE
•
Additional Filter Slide (4Filter)
•
Additional Filter Slide (6Filter)
•
Barcode Upgrade Kit