5. Maintenance
& Cleaning
2021-07
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
45
5.4
Cleaning the Instrument
WARNING
PRIOR TO CLEANING DISCONNECT THE INSTRUMENT FROM THE
MAIN POWER SUPPLY TO AVOID ANY RISK OF FIRE AND
EXPLOSION.
WARNING
THE CLEANING PROCEDURE SHOULD BE PERFORMED IN A
WELL-VENTILATED ROOM BY AUTHORIZED TRAINED PERSONNEL
WEARING DISPOSABLE GLOVES AND PROTECTIVE GLASSES AND
CLOTHING.
5.4.1
Cleaning the Device
Clean the outside of the device and the plate transport only with a dry or moist
cloth. If very dirty, clean it with a cloth moistened with a maximum of 70% ethanol
or mild detergent. Wipe dry with a lint-free cloth.
5.4.2
Liquid Spills
If any liquid is spilled on the instrument, it should be immediately removed to
prevent liquid running into the optical system causing loss of performance or the
error message
Lamp Low
due to one or more of the diode lenses not being
clean. Contact your local service representative in order to clean and check the
instrument.
WARNING
IF LIQUID SPILLED ON THE PLATE SUPPORT IS POTENTIALLY
INFECTIOUS IT SHOULD BE DISINFECTED ACCORDING TO THE
RELEVANT NATIONAL LAWS AND REGULATIONS.
Note
For disinfection solutions and procedure see chapter 5.6.1 and 5.6.2.