6. Quality
Control
2021-07
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
59
Note
Data can vary due to pipetting inaccuracy, meniscus of the liquid and the
application used.
Note
The limits of the instrument’s linearity will be measured with the QC plate.
Example for a tuneable wavelength instrument
Readings 0.100 to 2.500 OD
Dilution Factor
Absorbance
1
2.621 OD
0.5
1.323 OD
0.25
0.679 OD
0.125
0.360 OD
0.0625
0.192 OD
0.03125
0.110 OD
0
0.025 OD
Equation for linear regression: y = 2.5888x + 0.0314
Residual squared value: R
2
= 1