2. General
18
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
2021-07
2.3.4
SUNRISE Instrument Configuration with Tuneable
Wavelength Option (Gradient Filter Carriage)
PARAMETERS
CHARACTERISTICS
Measurement time:
Dual wavelength
Single wavelength
16 seconds
8 seconds
Wavelength range:
Gradient filter
Standard filter
400 - 700 nm
340 - 399 nm & 700 - 750 nm
Measurement range:
340 - 399 nm
400 - 750 nm
0 - 3.000 OD
0 - 4.000 OD
Resolution:
0.001 OD
Accuracy:
(492 nm)
0.000 - 2.000 OD
< (1.5 % + 0.010 OD) *
Precision:
(492 nm)
0.000 - 2.500 OD
< (1.0 % + 0.005 OD) *
Linearity:
(492 nm)
0.000 - 2.500 OD
< 2 %, R² >= 0.999
Wavelength selection:
Gradient filter
Special gradient filter any wavelength
between 400 and 700 nm in 1 nm steps.
The instrument can use up to eight
different filter carriages
* better than x % of measurement value plus corresponding OD value
Note: All deviations from the measurement value are meant in positive and negative direction.
Filter wavelength accuracy
Central wave/- 2 nm
Filter bandwidth
At 50% transmission:
450 nm
550 nm
650 nm
8.5 - 16 nm
10 - 15 nm
10 - 18 nm
Light source:
Halogen lamp 20 W
All connected devices must be approved and listed as per IEC 60950-1
Information Technology Equipment
– Safety and equivalent local standards.
Computer interface:
Serial RS 232 C
300 - 38,400 baud