3 - 36
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - Software Communication
Operating Commands
v <n>
Set Start Speed
The [v] command sets the speed at which the plunger begins its movement, in
pulses/sec. The plunger will then ramp up (slope) to the top speed. The start
speed should always be less than the top speed.
V <n>
Set Top Speed
The [V] command sets the top speed in pulses/second. This command may
be sent while a command string is already executing. (See section on
Changing Speed on the Fly, earlier in this chapter.)
Note: Syringes 2.5 mL and larger may require slower speeds. Users must
determine the appropriate speeds for their applications.
8
20000
9
22500
10
25000
11
27500
12
30000
13
32500
14
35000
15
37500
16
40000
17
42500
18
45000
19
47500
20
50000
Command
<n>
Parameter
Value
Default
Value
Description
v
50-1000
900
Set start speed in pulses/sec
Command
<n>
Parameter
Value
Default
Value
Description
V
5-6000
900
Set top speed in pulses/sec