3 - Software Communication
Error Codes and Pump Status
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - 51
Error Reporting Examples
0 1 X 0 0 1 1 1
47h
67h
71
103
7
Device not Initialized
0 1 X 0 1 0 0 0
48h
68h
72
104
8
Internal
failure
0 1 X 0 1 0 0 1
49h
69h
73
105
9
Plunger Overload
0 1 X 0 1 0 1 0
4Ah
6Ah
74
106
10
Valve Overload
0 1 X 0 1 0 1 1
4Bh
6Bh
75
107
11
Plunger Move Not Allowed
0 1 X 0 1 1 0 0
4Ch
6Ch
76
108
12
Internal Failure
0 1 X 0 1 1 1 0
4Eh
6Eh
78
110
14
A/D converter failure
0 1 X 0 1 1 1 1
4Fh
6Fh
79
111
15
Command Overflow
Status Byte
Hex # if Bit 5 =
Dec # if Bit 5 =
Error Code
[A7000R]
Does not move the plunger and reports a “No Error” status;
when queried ([Q] command), returns error. A
second try returns error 3 (67).
[P6000P600R]
Moves to position 6000, then stops. A [Q] command returns
an error.
[t2000R]
Returns an invalid command error immediately. The pump
status is “Not Busy”.
[A6000t2000R]
Returns an invalid command error immediately. The pump is
“Not Busy.”
Valve in Bypass
[A1000R]
Returns an error immediately; when queried ([Q] command),
does not return an error.