3 - 32
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - Software Communication
Operating Commands
Figure 3-7
Valve Position Examples for 9-Port Distribution Valves
3.5.2
Plunger Movement Commands
A <n> Absolute Position
The [A] command moves the plunger to the absolute position <n>, where
<n> = 0..6000 in standard mode and 0..48000 in fine positioning and
microstep mode.
For example:
–
[A300] moves the syringe plunger to position 300.
–
[A6000] moves the syringe plunger to position 6000.
Command
<n>
Parameter
Value
Description
A
0-6000
Absolute position in half increments (N=0)
0-48000
Absolute position in microsteps (N=1)
0-48000
Absolute position in microsteps (N=2)