3 - 2
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - Software Communication
XLP 6000 Addressing Scheme
Table 3-2
Address Switch Settings in Hex (ASCII)
The user can communicate with all pumps in the chain by using address “5Fh,” for
example to initialize all pumps at once. Then each pump can be controlled
independently by using addresses “31h” to “3Fh.”
Note: Multiple address commands cannot be used to determine device status or
to request reports. Each device must be queried separately to gather status or
generate a report.
Switch
Setting
Single Device
Dual Device
Quad Device
All Devices
Hex
Address
ASCII
Address
Hex
Address
ASCII
Address
Hex
Address
ASCII
Address
Address
Value to
Send
0
31
1
41
A
51
Q
5F
–
1
32
2
2
33
3
43
C
3
34
4
4
35
5
45
E
55
U
5
36
6
6
37
7
47
G
7
38
8
8
39
9
49
I
59
Y
9
3A
:
A
3B
;
4B
K
B
3C
<
C
3D
=
4D
M
5D
]
D
3E
>
E
3F
?
4F
O
F
Self Test