Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - Software Communication
Using the XLP 6000 Command Set
K<n>
Backlash Increments
The [K] command sets the number of backlash increments. The syntax for
this command is:
[K<n>]
where <n> = 0..31 in full step mode (12 is the default),
and <n> = 0..248 in fine positioning mode (96 is the default).
When the syringe drive motor reverses direction, the carriage will not move
until the backlash due to mechanical play within the system is compensated.
To provide this compensation, during aspiration, the plunger moves down
additional increments, then backs up the set number of backlash increments.
This ensures that the plunger is in the correct position to begin a dispense
move. Note that a small volume of fluid flows out the "input" side of the valve
during this operation.
>
Set User Data Command
The [>] command loads a byte of user data into non-volatile memory:
[> <n1>, <n2>], where: <n1> is 0..15 (location in non-volatile memory) and
<n2> is 0..255 (data to load into non-volatile memory).
U<n>
Write Pump Configuration to Non-Volatile Memory
The [U] command is used to write configuration information to the non-volatile
memory. The pumps are configured during the manufacturing process but
can be reconfigured at any time with the following [U] commands:
Table 3-5
Write Pump Configuration Command Values
Value<n>
Description
0
No Valve
1
3-Port valve
2
4-Port valve
3
3-Port distribution valve (face seal)
5
T Valve
7
6-port distribution valve
8
9-port distribution valve
9
Dual loop valve
11
3-Port distribution valve (plug)
30
Set Non-Volatile Memory Auto Mode
31
Clear Non-Volatile Memory Auto Mode
41
Set RS-232/RS-485 Baud rate to 9600