3 - 26
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - Software Communication
Initialization
W<n
1
>
Initialize Plunger Drive
The [W] command initializes the plunger drive only (commonly used for
valveless pumps). Because the valve is not initialized, only plunger force and/
or speed can be set.The default initialization speed is 500 pulses per second.
n
1
= Set initialization plunger force/speed
The parameters are described below.
<n
2
>
0
Sets initialization input port to port 1 (default)
1..X
Sets initialization input port for distribution valves,
where X is the number of ports on the valve.
<n
3
>
0
Sets initialization output port to port X (default), where X
is the number of ports on the valve.
1..X
Sets initialization output port for distribution valves,
where X is the number of ports on the valve.
W Parameter
Value
Description
<n
1
>
0
Initializes at full plunger force and at default initialization
speed (default)
1
Initializes at half plunger force and at default initializa-
tion speed
2
Initializes at one-third plunger force and at default ini-
tialization speed
10-40
Initializes at full force and at speed code <n
1
>. See
command <S> for a list of speed codes.