G - 8
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
G - Command Quick Reference
Error Codes and Status Byte
G.10 Error Codes and Status Byte
G.11 DA-15 Connector Pin Assignments
Status Byte
Hex # if Bit 5 =
Dec # if Bit 5 =
Error Code
7 6 5 4 3 2 1 0
0
or 1
0
or 1
Number
Description
0 1 X 0 0 0 0 0
40h
60h
64
96
0
No error
0 1 X 0 0 0 0 1
41h
61h
65
97
1
Initialization
0 1 X 0 0 0 1 0
42h
62h
66
98
2
Invalid command
0 1 X 0 0 0 1 1
43h
63h
67
99
3
Invalid operand
0 1 X 0 0 1 1 0
46h
66h
70
102
6
EEPROM failure
0 1 X 0 0 1 1 1
47h
67h
71
103
7
Device not initialized
0 1 X 0 1 0 0 1
49h
69h
73
105
9
Plunger overload
0 1 X 0 1 0 1 0
4Ah
6Ah
74
106
10
Valve overload
0 1 X 0 1 0 1 1
4Bh
6Bh
75
107
11
Plunger move not
allowed
0 1 X 0 1 1 0 0
4Ch
6Ch
76
108
12
Internal failure
0 1 X 0 1 1 1 0
4Eh
6Eh
78
110
14
A/D converter failure
0 1 X 0 1 1 1 1
4Fh
6Fh
79
111
15
Command overflow
Pin
Function
Remarks
1
24 V DC
2
RS-232 TxD line
Output data
3
RS-232 RxD line
Input data
4
Unused
5
CAN high signal line
6
CAN low signal line
7
Auxiliary input #1
TTL level
8
Auxiliary input #2
TTL level
9
Ground
Power and logic