2 - 4
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
2 - Hardware Setup
Communication Interfaces
Figure 2-1 shows the pin positions of the DA-15 connector on the printed circuit
assembly. This is a male connector that requires a female connector on the
mating cable.
Figure 2-1
DA-15 Connector Pins
2.3
Communication Interfaces
The computer or controller communicates with the XLP 6000 through an RS-232
interface, RS-485 interface, or CAN (Controller Area Network) interface.The XLP
6000 automatically detects the communication interface.
12
RS-485 B line
Data -
13
Auxiliary output #1
TTL level
14
Auxiliary output #2
TTL level
15
Auxiliary output #3
TTL level
Pin
Function
Remarks