Repeat Mode Settings
4-21
Repeat Mode Settings
To specify trigger action settings, from the
Setup
menu choose
Trigger
.
The settings in the above dialog box use a level trigger to start recording and an external trigger to stop
recording. A 10-second pre-trigger is specified, and recording is done once.
Pre/Post
Select this to enable a pre-trigger or post-trigger. When Pre/Post is selected but the lower
Repeat
is not selected, the
pre-trigger is enabled for starting recording by operations that use the toolbar or the buttons on the LX main unit. (The
post-trigger is not enabled for stopping recording by operations that use the toolbar or the buttons on the LX main unit.)
Start PreTrig/Stop PostTrig
Specify the length of the pre-trigger or post-trigger in seconds or in scan units. To switch the units, click a button on the
right. When 0 is entered, the pre- or post-trigger is disabled. Specify a value that does not exceed the time possible
using available free memory.
Switch between seconds and scans.
Enable pre- and post-triggers.
Specify the length of the pre-trigger.
Use a level trigger to start recording.
Enable the repeat mode.
Use an external trigger to
stop recording.
Number of repetitions.
Set the level trigger (next
page).
Set the recording time.
Содержание LX Series
Страница 2: ......
Страница 10: ...Contents viii...
Страница 26: ...Sampling Frequency and Number of Channels 1 16...
Страница 117: ...Main Unit Specifications 6 3 External Dimensions...
Страница 142: ...Recording Synchronization Specifications 6 28...
Страница 143: ...7 1 Section 7 Appendixes Troubleshooting 7 2 Supplied Accessories and Options 7 3 Supplied Accessories 7 3 Options 7 3...
Страница 146: ...Note 7 4 Note...
Страница 147: ......