Input Amp Settings
4-7
When using circuit signals (DC) of operations powered by an AC power source:
Input Mode: Unbalanced (Unbal) mode
If the frame ground (FG) of the input signal source is connected to ground, do not connect the LX frame ground
(FG) to ground.
When using circuit signals (DC) of operations powered by an AC power source:
Input Mode: Unbalanced (Unbal) mode
If there is no frame ground (FG) of the input signal source, connect the LX-10 frame ground (FG) to ground.
GND
FG
Vs
-
+
+
-
LX Main Unit
AR-LXPA
AR-LXPA
GND
FG
Vs
-
+
+
-
LX Main Unit