Input Amp Settings
4-9
ST Amp AR-LXST1
<ST1-8> is shown at the
Slot
tab for the AR-LXST1.
The AR-LXST1 is an input amp applied for strain gauges and strain gauge type sensors and DC inputs.
See Section 6 for the specifications.
Cautions on Using ST Amp
z
Calibration
It is suggested to use the LX ST amp after turning on the power for 5 minutes or more, and then executing
manual calibration.
In order to execute the manual calibration for precisely measurement, click
button on the toolbar of
the LX Navi. Also it is recommended to use the balanced input mode for precisely measurement.
z
Input Connection and Grounding
When using sensors:
Input Mode: Balanced (Bal) mode
Whenever possible, connect the LX frame ground (FG) to ground.
When using circuit signals (DC) of operations powered by an AC power source:
Input Mode: Balanced (Bal) mode
Always connect the frame ground (FG) of the input signal source and the LX frame ground (FG) to ground.
GND
FG
Vs
-
+
+
-
LX Main Unit
AR-LXST
GND
FG
Vs
-
+
+
-
LX Main Unit
AR-LXST
GND