Heating rate adjustment is particularly critical if
the transitions of an overlapped group are very
closely spaced in temperature. At 50
°
C/minute
the normal temperature lags in the TGA furnace
can be enough to overshoot the first transition,
which may reduce separation from the other
transitions in the group. This will be most
noticeable when the percent/minute baseline
preceding a rapid transition is relatively constant
and several orders of magnitude below the peak
rate of the transition. Running at 20
°
C/minute
instead of 50
°
C/minute is recommended for most
routine work using positive resolution settings
particularly if only one scan can be made of
each unknown material. Less than 10
°
C/minute
is usually not required.
For negative resolution settings the selection of
heating rates is usually quite different because
reaction rate overshoot must be minimized.
Generally, 1, 2, 5 and 10
°
C/minute are the most
useful rates to use with negative resolution
settings. Try 5
°
C/minute to start.
An important difference between the dynamic
rate (positive resolution setting) and constant
reaction rate (negative resolution setting) Hi-
Res
TM
mode, is that the maximum heating rate
selected is only an upper limit in negative mode,
but it is a gain factor in positive mode. In other
words, in negative mode the maximum heating
rate is irrelevant to heating control, except when
the reaction rate drops to such a low level that
heating rates higher than this maximum limit are
required to maintain the selected percent/minute.
In negative mode, as long as the maximum
heating rate is high enough to allow completion
of the transition at the selected percent/minute
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...