TA I
NSTRUMENTS
TGA 2950
B–51
Hi-Res
TM
Option
Abort Segment
The abort segment provides a mechanism to skip
over or terminate other method segments when
specific weight change conditions are met. This
segment has the following format:
Abort next seg if <signal> <condition>
<value>
where:
<signal> is the real-time measurement used to
decide whether to abort or not (weight % or
weight %/minute)
<condition> is the limit condition for aborting
(“<“ or “>”)
<value> is the abort limit which is compared
with the real-time signal
example:
Abort next seg if % < 20.0
Abort segments execute immediately when
encountered in a method and simply establish
conditions for testing the next segment. The
specified limit, designated by the
<signal>
,
<condition>
and
<value>
parameters, is tested
before and during the execution of the next
method segment. If the limit is reached at the
beginning of a segment, that segment is skipped
and method execution continues with the next
segment. If the segment following the abort
segment is equilibrate, initial temperature, ramp,
Hi-Res
TM
ramp, isothermal or step, and the limit
has not been reached yet, the limit will be tested
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...