TA I
NSTRUMENTS
TGA 2950
B–71
Hi-Res
TM
Option
Plastic
Laboratory Tubing
In Figures B.21 through B.25 we analyzed a
sample of Tygon
®
R-3603 plastic tubing, a poly-
vinyl chloride (PVC) based clear flexible tubing
commonly used in laboratories and industry. A
number of overlapped transitions are evident
between 100
°
C and 350
°
C, followed by two well
resolved transitions at approximately 400
°
C and
500
°
C. Of interest are the changes in resolution
of the various transitions as maximum heating
rate changes in both the conventional and Hi-
Res
TM
scans.
Observing the two conventional scans (Figures
B.21 and B.22) we see that the resolution of the
initial overlapped transitions is good in the 1
°
C/
minute scan, but poor in the 20
°
C/minute scan.
In contrast, the resolution of the two transitions
at 400
°
C and 500
°
C are best in the 20
°
C/minute
scan and reduced in the 1
°
C/minute scan. In
both cases, however, the small backside transi-
tion at 25% to 35% weight loss is barely
discernable. Comparing the conventional TGA
results in figures B.21 and B.22 to the 50
°
C/
minute Hi-Res
TM
scan in Figure B.23, we see
that all of these transitions are better resolved in
the Hi-Res
TM
TGA scan in a timely fashion.
Using Figures B.23, B.24, and B.25 we can
compare the result of changing the maximum
heating rate of a dynamic rate Hi-Res
TM
scan.
Maximum heating rates of 50
°
C/minute (Figure
B.23), 20
°
C/minute (Figure B.24) and 10
°
C/
minute (Figure B.25) were used with a method
consisting of a single Hi-Res
TM
ramp segment.
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...