Appendix B
B–32
TA I
NSTRUMENTS
TGA 2950
Thermocouple
Placement
Generally, the thermocouple should be placed 2
to 5 mm above the bottom of the sample con-
tainer (1 to 4 mm above the edge). Placement
1 mm above the edge is recommended for most
applications. Longer distances will not degrade
heater control and are sometimes helpful in
reducing any adverse effects of sample self-
heating or bubble formation.
Data Analysis
Effects
Hi-Res
TM
TGA, due to its technique of changing
heating rates dynamically, causes very signifi-
cant non-linearity of data points across transition
regions when plotted versus temperature. This
is not a problem when making plots, but when
automatic limit selection is used during data
analysis report generation the analysis program
can sometimes become confused about where to
place the limits. This situation can be identified
by curve tick marks and tangent lines which are
poorly placed on the weight curve or may even
be plotted off of the curve.
The difficulty is caused by the fixed size of the
transition onset and endset windows (the first
and last 12.5% of the analysis region). The data
analysis program assumes that data points are
uniformly distributed throughout this region. If
the number of data points in the baseline portion
of a window is very small compared to the
number of points in the sloped portion of the
window (the usual case with Hi-Res
TM
scans)
then the tangent lines will be excessively
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...