TA I
NSTRUMENTS
TGA 2950
2–15
GPIB Cable
1. Locate the GPIB connector on the right rear
of the TGA instrument (see Figure 2.6).
2. Connect the GPIB cable to the connector.
The GPIB cable is the only cable that fits
into the connector.
3. Tighten the hold-down screws on the
connector.
4. Connect the other end of the GPIB cable to
the controller or to the GPIB cable of
another TA Instruments instrument con-
nected to the controller. (For more informa-
tion, see your controller manual.)
5. Select an address from 1 to 9. Then use the
binary address switches on the TGA connec-
tor panel to set the desired address (See
Table 2.1). Figure 2.8 on the next page
shows a instrument address of 7.
If you have a multi-instrument system, each
instrument must have a different a different
address.
If you change the address after the TGA is
powered on, you must press the TGA’s
Reset button to enter the new address. Wait
until the instrument completes its startup
displays, then reconfigure the instrument
with the controller to bring the instrument
back online.
The instrument’s GPIB address is displayed during
start up and can also be viewed on the
instrument’s status display.
NOTE:
Installing the Instrument
NOTE:
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...