Adjusting Sensitivity
in Constant Reaction
Rate Mode
In constant reaction rate Hi-Res
TM
mode (nega-
tive resolution settings) the sensitivity setting is
used to adjust the heater control system to
minimize transition temperature overshoot and
heating control fluctuation. Higher sensitivity
settings result in decreased percent/minute
overshoot and tighter control at the beginning of
transitions. Lower settings have the opposite
effect.
In this mode sensitivity setting is used to adjust
the response of the heater control system to
changes in the rate of weight change (%/
minute). For materials that react gradually, low
sensitivity settings are generally preferred
because they help dampen noise and greatly
reduce the possibility of control cycling. How-
ever, when it is very important to avoid percent/
minute overshoot, or if the sample is highly
reactive then higher settings will be required.
The problem with too high a sensitivity setting
is that control cycling or heating rate ringing
may occur (see Figure B.11).
The recommended procedure to adjust sensitiv-
ity setting for constant reaction rate mode is to
start with a sensitivity setting of 1.0 and observe
the transition for percent/minute overshoot and
control cycling. If overshoot is acceptable, no
further adjustment is needed. If overshoot is
excessive, increase sensitivity setting by 1.0 and
recheck overshoot and cycling. Continue
increasing sensitivity until the results are
acceptable or until control cycling becomes
excessive.
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...