Appendix B
B–46
TA I
NSTRUMENTS
TGA 2950
this range. Setting 1.0 allows use of the full
range. Setting 2.0 reduces the range to about
one half the full range. Setting 3.0 to about a
third and so on up to 8.0. In general, higher
sensitivity settings bring the furnace to the
transition temperature more quickly but then
tend to remain at that temperature longer. In
other words, higher sensitivity settings bring the
control closer to stepwise isothermal heating.
(See Stepwise Isothermal Heating in the
section on using Abort Segments for more
information.)
The recommended procedure for adjusting
sensitivity setting for dynamic rate mode is to
start with a sensitivity setting of 1.0 and adjust
resolution setting to get the best separation
possible in the desired time frame. Then
increase sensitivity to 2.0, 4.0 and 8.0 to see if a
useful improvement in resolution results.
It is possible that no improvement in resolution
will result. This is usually caused by overlapped
transitions which are weakly temperature-
dependant and strongly time-dependant. In this
case, it doesnt matter how precisely we control
at a specific temperature, the individual compo-
nents of the sample material are all going to
decompose more or less together in the neigh-
borhood of the decomposition temperatures we
have selected via the resolution setting. About
all we can do is lengthen or shorten the total
time of decomposition by selecting larger or
smaller resolution settings.
The effect of changing sensitivity settings in
dynamic rate mode can be seen in Figure B.14
on page B-62.
Содержание TGA 2950
Страница 20: ...TA INSTRUMENTS TGA 2950 xx ...
Страница 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Страница 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Страница 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Страница 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Страница 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Страница 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Страница 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Страница 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Страница 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Страница 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Страница 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Страница 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Страница 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Страница 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Страница 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Страница 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Страница 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Страница 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Страница 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Страница 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...