deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 65/139
www.systec-controls.de
∫
=
dt
Q
Q
&
Calculating heat (quantity)
The heat quantity is derived as a function of thermal output over time.
]
/
,
[
h
kW
J
Q
=
8.7.2. Installing the PT100
The PT100 temperature sensors can be mounted on your piping using a
metal strap (photo), mounting chains, or a fabric-reinforced tape strap
(which is standard when buying deltawaveC-P package). Install the PT100
on the pipe as follows, for example:
Picture 35 Mounted temperature sensor (PT100)
The measuring method deployed for the PT100 is a relative
measurement. This means that the measured temperature values
with absolute reference may deviate from this measuring
equipment (e.g. compared to submersion thermometers). It is of
importance to set up a relation between both PT100s. The ideal
differential temperature between the PT100 temperature sensors
should amount to zero degrees prior to installation on the piping