deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 126/139
www.systec-controls.de
14.1.1.6.
Activate / Deactivate Auto-Window
When being in main menu (flow1) please choose OSZ to enter the
oscilloscope menu.
You now have to choose if you want to use auto window or not. In case of
problems with your measurement we recommend to try without auto
window first.
Choose „NO“
You should see distinguished signals (UP and DOWN) in the oscilloscope
as described in previous chapter.
If you want to activate auto-window please press DIAG and OSZ again.
Answer with YES. Please check if position of ultrasonic signals have been
changed. Please also check if value for delay has been changed. If yes,
please use auto window for your measurement (once you choose YES auto
window is activated also when you leave oscilloscope mode. To de-activate
auto window you have to enter oscilloscope mode again and answer with
NO)