deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 107/139
www.systec-controls.de
13.1.
Integrated sensor test function
Your deltawaveC offers the possibility to check the ultrasonic transducers in
combination with sensor cables.
For example: If there are no measurement results during a measurement than a
sensor test makes sure that the ultrasonic transducers works in a proper way. Thus
a non possible measurement is caused not by the instrument but possibly from the
application.
To perform a sensor test, following described steps are necessary:
1. Open OSCI window.
It is possible to open the OSCI window from the “DIAG” window or one of the three
measurement windows.
From that screens the OSCI window is reached by the press on the “OSCI” button.
A prompt with the question to apply autowindow function is shown. To perform a
sensor test it is necessary to proceed with “no”.
2. To get to the send signal the arrow button on the right side of the window (1) has
to be pressed several times until the displayed delay time (2) indicate “0.0”.
At the beginning of the y- line there are two signal packages (3). That packages are
the both send signals.
Picture 41 Osci window send signal
4. Proceeding with the test procedure there must be a little acoustic coupling grease
on one of the two transducers.
1
2
3