8.6. Zero Setting
It is advisable to run a zero calibration before you start measurements if
possible.
Prerequisite for error-free zero calibration is the complete
parameterization of the device, proper installation of both
ultrasonic transducers on the pipe, as well as their
electrical interconnection with the flow transmitter. Also
there should be really “zero-flow”. It is recommendable to
wait some minutes after stopping process / shut-off pipe to
allow the flow to calm down.
8.6.1. Zero calibration starting in "Flow1” menu
•
Close the valves of the piping.
•
Navigate to the "Setup" window as follows, using either of
three options:
-
After power on:
Select "Setup" within the start sequence
-
In the primary measuring window "Flow 1":
Select "Setup" and
“Zero Setup”
- Select "Set Zero"
The following window opens on completion of zero calibration: