deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 52/139
www.systec-controls.de
8.4.6. Mounting the ultrasonic transducers based on the Z method
Use a plastic or paper template to mark the mounting positions. The
example shows how to mark the positions using a plastic template.
•
Wrap the plastic template once around the pipe at the mounting
position of the first ultrasonic transducer (transducers face has to
be in line with the line to be drawn).
•
Using a felt tip pen, draw a line on the pipe along the template
(corresponds with the pipe circumference)
Picture 26 Attaching plastic template
Picture 27 Attaching plastic template 2
Draw line