deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 127/139
www.systec-controls.de
14.1.2. Diagnostic menu
The diagnostic menu shows a lot of parameters which are
mainly readable for experienced users. The diagnostic data are
also suited for troubleshooting
Diagnostic window 1 / 3:
Frequency:
Shows the signal frequency which depends on the selected transducer
Basically there are three different transducer using different frequencies
(500kHz, 1000kHz, 2000kHz)
Signal:
Shows the type of signal code. deltawaveC-P uses coded signals to allow
reliable detection of signals.
T1:
Signal run time signal 1 (Upstream signal or downstream signal depending
on connection of cables).
T2:
Signal run time of back signal 2 (Upstream signal or downstream signal
depending on connection of cables).
dT:
Measured difference between T1 and T2 (minus zero offset if applicable).
dT is proportional to flow.
vF:
Measured velocity
Vs: