systec Controls Mess- und Regeltechnik GmbH 114/139
www.systec-controls.de
signal is usually weak indicated by low amplitude. The V-mode signal is
usually stronger than the W-mode signal.
The time between the different signals might be significantly higher or lower
depending on pipe sizes and speed of sound of involved materials.
A different situation occurs when measuring in Z-mode
Picture 46 Signals in Z-mode
Beside the send signal only the desired signal (Z-signal) is visible. The Z-
signal reaches the receiving transducer after T=t1z
14.1.1.2.
Oscilloscope menu of deltawaveC-P
With the oscilloscope menu you can analyse signals and even manipulate
them in order to handle very challenging applications.