AMI Codes-II TC
Program List and Explanations
A-96.250.631 / 280219
91
5.3.1.1
Disinfection
5.3.1.1.1
Total chlorine 1
5.3.1.1.1.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E001 is displayed in the
message list. Range: 0.00–20.00 ppm
5.3.1.1.1.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0.00–20.00 ppm
5.3.1.1.1.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value
fluctuates around the alarm value.
Range. 0.00–20.00 ppm
5.3.1.1.1.45
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0.00–28‘800 sec
5.3.1.1.2
Total chlorine 2
5.3.1.1.2.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E003 is displayed in the
message list. Range: 0.00–20.00 ppm
5.3.1.1.2.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00–20.00 ppm
5.3.1.1.2.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value
fluctuates around the alarm value.
Range. 0.00–20.00 ppm
5.3.1.1.2.45
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0.00–28‘800 sec
5.3.1.1.3
Calc. dichl.
5.3.1.1.3.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E005 is displayed in the
message list. Range: 0.00–20.00 ppm
5.3.1.1.3.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E006 is displayed in the message
list. Range: 0.00–20.00 ppm