AMI Codes-II TC
Maintenance
A-96.250.631 / 280219
55
6.7.
Cleaning the Photometer
Clean the photometer after indication by alarm (E020, FOME dirty).
Switch off the instrument according to instructions in
Material
Small brush.
Procedure
1
Close the flow regulating valve [A].
2
Wait until the sample flow through the photometer has stopped.
3
Unscrew the cover [B] from the photometer [C].
4
Clean the Photometer with a small brush [D].
5
Screw the cover to the photometer.
6
Open the flow regulating valve.
A
B
C
D
Flow regulating valve
Photometer cover
Photometer
Brush
A
B
C
D
C