80
A-96.250.631 / 280219
AMI Codes-II TC
Program List and Explanations
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active,
acknowledged). If an active error is acknowledged, the alarm relay
is active again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5
Demands necessary maintenance, e.g. preparing new reagents.
1.2 Message List
1.3.1
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation
:
View the Designation of instrument.
Version
: Firmware of instrument (e.g. V6.20-05/18)
2.1.3
Peripherals
: PeriClip 1: Firmware of peristaltic pump (e.g. 1.06)
2.1.4
Factory Test
: Test date of the Instrument, Motherboard and
Frontend. QC factory test.
2.1.5
Operating Time
: Years/Days/Hours/Minutes/Seconds
2.2 Sensors
2.2.1
Photometer:
Absorbance
: Process value, depends on sample.
Raw value
: Shows the actual photometer signal in Hz.
2.2.1.4
Cal. History:
Shows the diagnostic values of the last calibrations.
Number
: Counter for the calibrations
Date, Time
: Date and time assigned to a number.
Slope
: Slope photometer: 0.8–1.2