AMI Codes-II TC
Product Description
A-96.250.631 / 280219
15
2.1.
Instrument Specification
Power Supply
Voltage:
Power consumption:
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
Electronics
Aluminium with a protection degree of IP 66 / NEMA 4X
housing
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
min. 10 l/h
5 - 50 °C (41 - 122 °F)
0.15 - 2 bar (2 - 28 PSI)
pressure free
NOTICE:
No oil, no grease, no sand.
On-site
The analyzer site must permit connections to:
requirements
Sample inlet:
Sample outlet:
Tube 6 x 8 mm
2 Drains, 1/2” hose nozzle for flexible
tube diam. each 20x15 mm