AMI Codes-II TC
Instrument Setup
A-96.250.631 / 280219
37
4.3.
Establish Sample Flow
WARNING
Water pollution
The drain of the photometer outlet contains DPD.
At no means recirculate it into the water system.
1
Open the flow regulating valve (D) and wait until the flow cell is
completely filled.
2
Switch on power.
3
Adjust the sample flow so that always a small part of the sample
drains off through the overflow tube.
4
Start <Fill system>, see
Fill or Flush Reagent System, p. 38
A
B
C
D
E
F
Cover
Constant head tube
Flow cell block
Flow regulating valve
Filter
Filter vessel
Level
A
B
C
D
E
F