AMI Codes-II TC
Maintenance
A-96.250.631 / 280219
49
6.3.1
Reagents for Measuring Total Chlorine
Reagents needed:
1 x 50 ml of Oxycon On-line DPD
1 x 240 gr of Oxycon On-line Buffer
1 x 60 gr of Reagent Oxycon On-line KI
4 liters demineralized water
Prepare
Oxycon
On-line DPD
1
Rinse the canister [E] labelled OXYCON ON LINE DPD-Re-
agent with demineralized water.
2
Fill the canister up to the 2 liter mark with demineralized water.
3
Slowly pour the content of one bottle of concentrate Oxycon
On-line DPD (50 ml) into the canister.
Avoid splashing!
4
Close the canister with the screw cover and tighten it well.
5
Mix the demineralized water and the reagents well.
6
Put the canister [E] into the holder [G].
7
Remove the screw cover, insert suction lance [A] and tighten
the screw cover.
Prepare
Oxycon
On-line Buffer
& KI
1
Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
tion with demineralized water.
2
Fill the canister up to the 2 liter mark with demineralized water.
3
Slowly pour the content of one bag of Oxycon On-line Buffer
into the canister.
Avoid splashing!
4
Add the content of one bag Oxycon On-line KI to the same can-
ister.
5
Close the canister with the screw cover and tighten it well.
6
Mix the demineralized water and the reagents well.
7
Put the canister [F] into the holder [G].
8
Remove the screw cover and insert suction lance [B]. Tighten
the screw cover.
9
Fill reagent system. See