96
A-96.250.631 / 280219
AMI Codes-II TC
Program List and Explanations
5.3.2.24
Mode:
Operating mode (interval, daily, weekly)
5.3.2.340
Interval:
The interval can be programmed within a range
of 1–1’440 min.
5.3.2.44
Run Time
: Enter the time the relay stays active.
Range: 5–32’400 sec.
5.3.2.54
Delay
: during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6’000 sec.
5.3.2.6
Signal Outputs
: Select operating mode of the signal output:
5.3.2.7
Output/Control
: Select operating mode of the controller output:
The relay contact can be activated daily, at any time of a day.
5.3.2.341
Start time
: to set the start time proceed as follows:
1
Press [Enter], to set the hours.
2
Set the hour with the [
] or [
] keys.
3
Press [Enter], to set the minutes.
4
Set the minutes with the [
] or [
] keys.
5
Press [Enter], to set the seconds.
6
Set the seconds with the [
] or [
] keys.
Range: 00:00:00–23:59:59
5.3.2.44
Run Time
: see Interval
5.3.2.54
Delay
: see Interval
5.3.2.6
Signal Outputs
: see Interval
5.3.2.24
Interval
Cont.:
Signal outputs continue to issue the measured value.
Hold:
Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Off:
Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
Cont.:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.
5.3.2.24
daily