confirm the calibration results. A measurement cannot be in
progress while the automatic calibration is being performed.
Note: SV 971A has two different dynamic ranges
– Normal and Low. In the SLM functions
(Level Meter, 1/1 Octave, 1/3 Octave, RT60 and STIPA) both ranges can be used for
measurement, but in the SEM functions (Dosimeter
,
1/1.&Dose
,
1/3.&Dose)
– only Normal
). But in case of calibration, it does not matter which function and
which range is selected.
In the
Simple
user interface, the
Calibration
list
comprises positions enabling calibration with the use of
the sound calibrator (
By Measurement
), checking and
erasing previous calibration records (
Last Calibration
,
Calibration History
,
Clear History
), adding current
calibration results to the logger file (
Post Calibration
).
<ENT>
In the
Advanced
user interface, there is additional position in the
Calibration
list
-
Auto Calibration
, which allows you to switch on/off the auto calibration
function.
Note: It is advised to perform calibration of the instrument each time before the measurements
begin. A single calibration at the start of each day is usually sufficient for most regulations.
Note: The calibration factor is always added to measurement results and measurement range
limits of all measurement functions.
Note: The recommended factory calibration interval is 12 months for instruments to be
confident in their continuing accuracy and compliance with the international codes. Please
contact your local Svantek distributor for further details.
Note:
It is possible to perform so called “By Sensitivity” calibration by setting the calibration
factor with the use of special command
– see Appendix A.
3.2.1 Calibration
– By Measurement
To calibrate the instrument:
1. Set the calibration level (
Cal. Level
)
– see Appendix C, Chapter C.1, par.
Calibration.
2. Attach the sound calibrator (SV 36 or equivalent 114 dB/1000 Hz)
carefully
over the microphone of the instrument.
Note: It is also possible to use an electro-mechanical pistonphone, which generates a signal
(ca 124
dB) or different type of acoustic calibrator dedicated for ½” microphones. It is also
necessary to switch the instrument Range to the Normal level.
3.
Switch on the calibrator (if the used calibrator doesn’t have auto run function)
and wait ca 30 seconds for the tone to stabilise before starting the calibration
measurement.
4. Start the calibration measurement by pressing the
<Enter>
or
<Start>
key.
The calibration delay time is set to 3 seconds. While waiting for the start of the
measurement the
Delay
is counting down on the display.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...