SV 971A User Manual
39
The
Logging
list enables programming of the logging
functions: recording of summary and logger results
(measurement history) in a logger file and recording of
audio signal in a wave file.
In the
Simple
instrument interface mode, the
Logging
list includes only of one position -
Logger Set.
<ENT>
4.5.1 Setting general logging parameters
– Logger Setup
The
Logger Set.
list enables activating the logging
function (
Logger
) and programming logger general
parameters.
The
Logger
position switches
On
or
Off
the logging
functionality.
►
Switching on the
Logger
(
On
) activates two positions in the
Logging
list, which
enable saving selected results from the three profiles and spectra
(
Logger Results
) with the step defined by the
Logger Step
parameter and
programming the
Logger Trigger
.
Note: If Logger is Off, result files are not created, and measurement results (both summary
and logger) are not saved.
Splitting logger files
The
Log.
Split
position enables splitting the data logging into separate files. If
Log.
Split
is
Off
the data
will be logged in one logger file with the name defined in the
Logger Name
position. Default value:
Off
.
In other cases, the registration will be carried out in separate files and the
registration in the new file will start after expiration of integration period
(
Integr. Per
), or at every quarter of the RTC (
Sync. to 15m
), or at every half an
hour of the RTC (
Sync. to 30m
), or at every hour of the RTC (
Sync. to 1h
), or at
the specified by the user times (
Spec. Time
). Whenever the split time is achieved
the logger file is closed and the new file with the increased by one number is
opened for subsequent measurement data.
Note: In the Simple interface mode, the Logger Split parameter is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Off)
.
If
Spec. Time
is selected in the
Log.
Split
position, you
can set six split times (
Split Time1
,
Split Time2
,
Split Time3
,
Split Time4
,
Split Time5
and
Split Time6
) changing
Off
to the desired time of the day
when splitting should occur.
►
The
Logger Step
defines the step for logger results logging in a file. It can be set
from
100ms
to
1h
. Its value by default is set to
1s
.
Note: For logger steps smaller than 1s, the running Leq results are calculated with a step of
1s but saved to the logger file with the logger step.
In such cases, the logger curve for the
running Leq results will be a stepped curve.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...