SV 971A User Manual
41
4.5.3 Configuring Logger trigger
– Logger Trigger
In the
Logger Trigger
screen, you can configure the
way the logger results are to be registered in the logger
file. It is a context list of parameters in which the trigger
can be switched
Off
or On by selecting its type (
Level+
or
Level
–
) in the
Trigger
position.
<ENT>
When the trigger is selected, other parameters appear in the list: a measured result that is checked for a
trigger condition (
Source
), a threshold level (
Level
) as well as a number of results saved in the logger
before the trigger condition is met (
Pre
) and the number of the results saved in the logger after the last
trigger condition is met during logging (
Post
).
Note: In the Simple interface mode, the Logger Trigger position is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Trigger: Off)
.
Level trigger
The
Level+
/
Level
–
trigger enables logging of the time-history results
(
Logger Results
) with the
Logger Step
under the condition: value of the LEQ
result (
Source
) averaged by the
Logger Step
period is greater/lower than the
threshold value (
Level
). In other cases, the logging is skipped. Due to this type
of trigger it is possible to separate results related to the low/high noise level.
The logging is active only when the
summary results are measured, i.e. from
the
measurement
start
till
the
measurement stop.
This means, for example, that when the
measurement is waiting for a trigger
condition, logging is skipped, even if the
logger trigger condition is met.
Note:
When logging is waiting for the level trigger the “level” icon
appears alternatively with the „logger” icon.
/
Source result
Only one measured result (
Source
) can be used for checking trigger condition in the
Level Meter
mode,
namely the instantaneous LEQ from the first profile (with appropriate filter and detector), which is denoted
here as
Leq(1)
. This position cannot be changed.
Threshold level
The threshold value (
Level
) can be set in the range from
24 dB
to
136 dB
. The
Source
value compares with the
Level
value every 0.5 milliseconds.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...