SV 971A User Manual
34
Delay of measurement start
The
Start Delay
parameter defines the delay period from the
<Start/Stop>
keystroke to the real start of the measurement (digital filters of the instrument
constantly analyse the input signal even when the measurement is stopped). This
delay period can be set from
0 second
to
60 minutes
. Its value by default is set
to
1s
.
Note: In the Simple interface mode, the Start Delay parameter is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (1s)
.
Note: The minimum delay period is equal to 0 second. In the Calibration mode, the delay
period is always equal to 3 seconds.
Note: After turning on, the instrument requires 30 seconds to warm up. If you press the
<Start/Stop> key earlier, the instrument will start a measurement not earlier than after 30
seconds.
Synchronisation of measurement cycles
The
Start Sync.
parameter defines synchronisation points with the instrument’s
RTC. The
Start Sync.
parameter can be set as:
Off
,
1m
,
15m
,
30m
and
1h
. For
example, if
1h
is selected, the measurement will start from the beginning of the
first second of next hour after the
<Start>
keystroke, and then will be repeated
also from the first second of the following hour after elapsing the integration
period if the number of cycles is greater than one. The default value is set to
Off
.
Integration period
The
Integr. Per
parameter defines the period during which the signal is being
measured (and for some results averaged/integrated) and measurement results
are logged in the logger file as
Summary Results
(see description of the
Logger Setup
). The integration period can be infinite (
Inf
) or selected from the
set:
24h
,
8h
,
1h
,
15m
,
5m
,
1m
, from
1s
to
59s
with 1s step, from
1m
to
59m
with
1m step, from
1h
to
24h
with 1h step.
During the Integration Period, the instrument performs series of 1-second measurements/integrations,
and every second averages 1-second results with the results averaged for the n-1 seconds. These
averaged results are displayed and renewed every second for the elapsed measurement time (n
seconds). In the end of the Integration Period the averaged measurement results are saved in the logger
file providing that such saving is switched on.
The measurement will stop automatically after this period and start again if the number of measurement
repetitions (
Rep. Cycles
) is greater than one.
The definitions of the measurement results in which the integration period is used are given in
Appendix D.
Number of measurement cycles
The
Rep. Cycles
parameter defines the number of measurements (with the
measurement period defined by the
Integr. Per
parameter) to be performed by
the instrument after the
<Start>
keystroke. The
Rep. Cycles
number values are
within the limits [Inf, 1÷1000]. Its value by default is set to
1
.
For example, if
Integr. Period
is equal to 8 hours and
Rep. Cycles
is equal to 2, the instrument performs
first integration for the 8-hour period from the measurement start and second integration for the 8-hour
period from the end of the first integration. At the end of each cycle the 8 hours LEQ will be saved in the
logger file.
Note: In the Simple interface mode, the Rep. Cycles parameter is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (1)
.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...