SV 971A User Manual
23
Help information
In most screens, the last line or two lines contain help information: how to select
or modify the parameter’s value, change the character in the text line etc. For
example,
Delete: Shift <
means that you can delete the selected position with
the
◄
key pressed with
<Shift>
.
Inactive parameters
If some functions or parameters are not available, the positions in the menu or
parameter lists linked with this function or parameter become inactive (the
selected line field will be in the frame with black background, not yellow). For
example, if
Logger
(
path: <Menu> / Measurement / Logging / Logger Set.)
is
switched off, some other
Logging
positions will be not
active!
2.6 D
EFAULT SETTINGS
Factory settings
The instrument as sold has default settings which you may change, but always
return to them with the use of the
Factory Settings
function of the
Auxiliary Setup
section.
Next chapters of the manual describe in detail what each parameter means and
how to change the instrument settings.
Main default settings
With default settings, the instrument has the
Simple
user interface and is
configured as the Sound Level Meter (
Measurement Function: Level Meter
) to
measure sound pressure level by three virtual meters, so called profiles, with
1 second delay from the
<Start>
key pressure, infinite integration time
(
Integration Period: Inf
) and linear Leq integration (
LEQ Integration: Linear
).
Sound pressure is measured with compensation of microphone internal noise
and case effect in the free field (
Microphone: On
,
Field Comp.: Free Field
,
Windscreen: On
), active logging of the selected results (
Lpeak
,
Lmax
,
Lmin
and
Leq
) with 1 second step for all profiles and summary results saving.
Other functions are switched off, like measurement trigger, logger trigger, event
recording and timer.
The logger and summary results will be automatically saved in the file with the
name presented in the
Logger Setup
list (
Logger Name: Lxxxx
). The logger
results will be logged with 1 s step (
Logger Step: 1s
). Logger splitting is
Off
.
Default Profile settings:
Profile 1
-
C
weighting filter for Peak results (
Filt.Peak(1)=C
),
A
weighting
filter for other results (
Filter(1)=A
),
Fast
for the LEQ detector
(
Detector(1)=Fast
);
Profile 2
-
C
weighting filter for Peak results (
Filt.Peak(2)=C
),
C
weighting
filter for other results (
Filter(2)=C
),
Fast
for the LEQ detector
(
Detector(2)=Fast
);
Profile 3
-
Z
weighting filter for Peak results (
Filt.Peak(3)=Z
),
Z
weighting
filter for other results (
Filter(3)=Z
),
Fast
for the LEQ detector
(
Detector(3)=Fast
);
You can change all above-mentioned settings using the
Profiles
position of the
Measurement
section.
The instrument remembers all changes by the next time it is used. You can return to default settings (set
up by the manufacturer) with the use of the
Factory Set.
position in the
Aux. Setup
section.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...