SV 971A User Manual
– Appendixes
142
Note: All bytes of that transmission are ASCII characters.
Note: Changing settings using #1 functions during measurements running state (#1,S1;) is
blocked. Stop the measurements (#1,S0;) before changing the settings.
In order to read out all current control settings the user should send to the device the following sequence of
characters:
#1;
In this case the instrument outputs all control settings given in A.3
#1,Xccc,Xccc,(...),Xccc;
Example:
The instrument sends the following sequence of characters as an answer for the mentioned above
request:
#1,U971,N1234,W1.05.1,Q0.01,M1,R1,F2:1,F3:2,F1:3,F2:4,F3:5,F1:6,J2:1,J3:2,J1:3,J3:4,J3:5,J1:6,f1,C
1:1,C0:2,C2:3,C1:4,C0:5,C2:6,B0:1,B3:2,B15:3,b0,d1s,D10s,K5,L0,Y3,y0,XT0,XL100,XQ0,Xq0,XC115:
1,XC115:2,XC115:3,Xl115:1,Xl115:2,Xl115:3,XA0,XD-1:1,XD-1:2,XD-1:3,XD-1:4,XD-1:5,XD-
1:6,S0,T1,e480,c1:1,c1:2,c1:3,h0:1,h0:2,h0:3,x3:1,x3:2,x5:3,m0,s0,l100,O10,o0,t0;
means that:
•
SV 971A is investigated (
U971
); see #7,US; command for unit subtype information;
•
its number is 1234 (N1234);
•
software version number is 1.05.1 (W1.05.1);
•
calibration factor is equal to 0.01 dB (Q0.01);
•
Level Meter
is selected as the measurement function (M1);
•
range is
Low
(R1);
•
A
filter is selected in profile 1, SLM function (F2:1);
•
C
filter is selected in profile 2, SLM function (F3:2);
•
Z
filter is chosen in profile 3, SLM function (F1:3);
•
A
filter is selected in profile 1, SEM function (F2:4);
•
C
filter is selected in profile 2, SEM function (F3:5);
•
Z
filter is chosen in profile 3, SEM function (F1:6);
•
A
Peak filter is selected in profile 1, left channel, SLM function (J2:1);
•
C
Peak filter is selected in profile 2, left channel, SLM function (J3:2);
•
Z
Peak filter is selected in profile 3, left channel, SLM function (J1:3);
•
C
Peak filter is selected in profile 1, both channels, SEM function (J3:4);
•
C
Peak filter is selected in profile 2, both channels, SEM function (J3:5);
•
Z
Peak filter is selected in profile 3, both channels, SEM function (J1:6);
•
Z
filter is selected for
1/1 Octave
or
1/3 Octave
analysis (f1)
•
Fast
detector is selected in profile 1, SLM function (C1:1);
•
Impulse
detector is chosen in profile 2, SLM function (C0:2);
•
Slow
detector is selected in profile 3, SLM function (C2:3);
•
Fast
detector is selected in profile 1, SEM function (C1:4);
•
Impulse
detector is chosen in profile 2, SEM function (C0:5);
•
Slow
detector is selected in profile 3, SEM function (C2:6);
•
logger’s buffer is not filled by the results from profile 1 (B0:1);
•
Lpeak
and
Lmax
values are stored in the files of the logger from profile 2 (B3:2);
•
Lpeak
,
Lmax
,
Lmin
and
Leq
values are stored in the files of the logger from profile 3 (B15:3);
•
results of
1/1 Octave
or
1/3 Octave
analysis are not stored in the files of the logger (b0);
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...