SV 971A User Manual
75
LEQ detector
Following LEQ detectors are available in the instrument:
Imp.
,
Fast
and
Slow
.
Dosimeter
specific parameters can be set in accordance with the OSHA HC
(Occupational Safety and Health Administration - Hearing Conversation),
OSHA PEL (Occupational Safety and Health Administration
– Permissible
Exposure Level) and ACGIH standards.
Criterion Level
is a
steady noise level permitted for a full eight-hour work shift
:
60dB
,
65dB
,
70dB
,
75dB
,
80dB
,
84dB
,
85dB
,
87dB
,
90dB
;
Threshold Level
is a
noise level limit below which the dosimeter does not
accumulate noise dose data
:
None
,
60dB
,
65dB
,
70dB
,
75dB
,
80dB
,
85dB
,
90dB
;
Exchange Rate
is an
amount by which the permitted sound level may increase
if the exposure time is halved
:
2
,
3
,
4
,
5
,
6
;
ULT Threshold Level
(Upper Level Time)
is a threshold level for calculation of
the ULT results:
70dB
÷
140dB
;
PTC Threshold Level
(Peak Threshold Counter) is a threshold level for
calculation of the PTC results:
70dB
÷
140dB
.
11.4 C
HECKING MEASUREMENT RANGE
–
R
ANGE
The measurement range for the
Dosimeter
function is
set to
Normal
and you cannot change it. The detailed
description of the measurement range parameters is
given in Appendix C.
<ENT>
11.5 S
ETTING EXPOSURE TIME
–
E
XPOSURE
T
IME
The
Exp. Time
position enables setting the desired
value of the workday exposure time which is used for
calculation of the
LEPd
results (see Appendix D).
<ENT>
11.6 S
ETTING ALARM THRESHOLDS FOR DOSE METER RESULTS
–
A
LARM
The
Alarm
position is active only in the
Dosimeter
function and enables programming the alarm thresholds
for three profiles (
Thresh. P1 (2,3)
).
Thresholds can be set for next measurement results of
the
Dosimeter
in ranges:
<ENT>
DOSE
: 1÷200%;
D_8h
: 1÷200%;
PTC
: 1÷1000;
ULT
: 1÷60s.
If
Off
is selected, the alarm for the measurement result
is switched off.
<ENT>
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...