Using and Controlling the AirSeal® iFS in its Different Modes
EN
2. At the conclusion of the initial self test, insert the Tri-Lumen Filtered Tube Set
and
AirSeal Mode
will automatically launch. Alternatively, press the
AirSeal
Mode
key to select this operating mode.
Inserting the Tri-Lumen Filtered Tube Set
•
To be carried out by non-sterile technician:
• Open packaging of the Tri-Lumen Filtered Tube Set.
• Have a sterile technician remove the tube set from the inside of the package.
To be carried out by sterile technician:
• Keep the distal end of the Tri-Lumen Filtered Tube Set (AirSeal® Cannula con-
nector) in the sterile area and hand the tube end with filter to the non-sterile
technician.
Option a) Initial insufflation with Veress needle or conventional cannula
(advised)
Use the Single Lumen Adapter to connect the tri-lumen tube with the Luer
lock connector of a conventional Veress needle or a conventional cannula
(see 5.5.2 "Insufflation Tube Sets"). The Single Lumen Adapter comes con-
nected to the tri-lumen tube upon delivery. Ensure the connection is tight
before use.
Option b) Initial insufflation with AirSeal® Access Port
Connect the Tri-Lumen Filtered Tube Set with the AirSeal® Access Port (see
6.1.1 "AirSeal® Access Port").
To be carried out by non-sterile technician:
• Insert the Tri-Lumen Filtered Tube Set into the receptacle at the front of the
device. Use the lever to lock the filter housing in place or to unlock and re-
lease it.
3. Insert the Veress needle, conventional trocar, or AirSeal® Access Port in accor-
dance with proper laparoscopic technique.
4. Choose desired settings.
Setting the nominal flow and nominal pressure (increase/decrease) is possi-
ble during insufflation or while insufflation is stopped.
Setting nominal flow:
The device features three different flow level rates that can be adjusted in the
user menu (see chapter 8 "User Menu" for additional information). The facto-
ry defaults are:
Level 1 -> 5 l/min
Level 2 -> 20 l/min
Level 3 -> 40 l/min
• To set the flow rate, press the key
(1)
,
(3)
or
(4)
.
Setting the nominal pressure:
Press the key
»
or
º
(
(5)
or
(8)
), to set the nominal pressure. Values may
range from
5 to max. 20 mmHg
and can be adjusted in increments of 1. The
starting value for the pressure can be set in the user menu within a range of
5 - 15 mmHg.
• Scrolling is enabled by keeping the
»
or
º
key depressed longer than 1.5
distal end
proximal end
(3)
(1)
(4)
(5)
(8)
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...