Using and Controlling the AirSeal® iFS in its Different Modes
27
EN
sterile technician.
• Connect the Luer lock connector of the clear tube to the conventional can-
nula. Open the inflow valve/stopcock.
• Follow prompts on iFS. When instructed, connect the Luer lock connector of
the blue tube to a different conventional cannula to activate
smoke evacu-
ation.
To be carried out by non-sterile technician:
• Insert the Smoke Evacuation filter housing into the receptacle at the front
of the device. Use the lever to lock the filter housing in place or to unlock
and release it.
3. Choose desired settings.
Setting the nominal flow and nominal pressure (increase/decrease) is possi-
ble during insufflation or while insufflation is stopped.
Setting nominal flow:
The device features three different flow level rates that can be adjusted in the
user menu (see chapter 8 "User Menu" for additional information). The facto-
ry defaults are:
Level 1 -> 5 l/min
Level 2 -> 20 l/min
Level 3 -> 40 l/min
• To set the flow rate, press the key
(1)
,
(3)
or
(4)
.
Setting the nominal pressure:
Press the key
»
or
º
(
(5)
or
(8)
) to set the nominal pressure. Values may
range from
5 to max. 20 mmHg
and can be adjusted in increments of 1. The
starting value for the pressure can be set in the user menu within a range of
5 - 15 mmHg.
• Scrolling is enabled by keeping the
»
or
º
key depressed longer than 1.5
seconds.
Increasing the nominal pressure to > 15 mmHg:
Status line depicts the message:
Safety threshold
:
Pressure >15 mmHg!
This is where the recommended range for intra-abdominal and colorectal
pressure ends. Release the key for 2 seconds, then pressure can be increased
above 15 mmHg.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
4. Start insufflation by pressing the
START
key.
The status line depicts
Initial Insufflation
.
The initial insufflation ends when
the nominal pressure value has been reached for the first time. The device
prompts the user to connect the blue tube to enable Smoke Evacuation func-
tion.
To enable smoke evacuation, connect the blue tube of the bifurcated tube set
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...