Device Purpose
EN
WARNING!
Lowest flow and pressure
Depending on age and health condition of the patient, the smallest possible flow
and pressure for establishing the pneumoperitoneum or pneumorectum should
be selected. It is not recommended to exceed insufflation pressures of 15mmHg
in colo-rectal procedures.
WARNING!
Laparoscopy with CO
2
should not be performed on children with cardiovascular
problems.
WARNING!
Dehydration
Insufflation can lead to dehydration of the tissue. This can result in organ tissue
damage and cardiovascular reactions of the patient. Long surgeries and large
leaks increase the risk of dehydration (especially at the insertion points of the
trocars or when changing instruments).
WARNING!
Embolism/insufflation of internal organs
Improper placement of the insufflation instrument could cause gas penetrating
a vessel or an internal organ, resulting in gas embolisms. To reduce the risk, use
a low flow rate for the first insufflation and ensure that the insufflation instru-
ment is correctly positioned. Check the position of the insufflation instrument
immediately if the actual pressure rapidly reaches the nominal pressure value.
gas embolisms can also be caused by a high intra-abdominal pressure. Avoid
high-pressure settings and close damaged blood vessels at once.
WARNING!
Emphysema
Incorrect placement of a cannula or a trocar into subcutaneous tissue may lead
to emphysema. To reduce the risk, use a low gas flow rate for the first insuffla-
tion and ensure that the insufflation instrument is correctly positioned. Long sur-
geries (> 200 min.), the use of many access points, duration and size of leaks at
these points may also contribute to emphysema. Be sure to close leakages in tro-
car access points immediately.
WARNING!
Additional insufflation sources/automatic venting system
Make sure the automatic venting system is activated (see chapter 8 "User
Menu"). The use of additional insufflation sources increases the intra-abdominal
pressure. Continuously monitor intra-abdominal pressure over the course of the
entire insufflation if additional sources are used.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...