Informational, Warning and Error Messages
47
EN
Error message
Cause
Remedy
Venting valve error! Restart the device.
If the error occurs again call service!
Valve malfunction.
Restart the device. If the error occurs
again, call service!
Electronic error! Restart the device. If
the error occurs again, call service!
Malfunctions in the electronic system.
Restart the device. If the error occurs
again, call service!
Sensor error! Restart the device. If the
error occurs again call service!
Sensor malfunction.
Restart the device. If the error occurs
again, call service!
Device temperature error! Do not use
device. Call service!
Temperature in device too high.
Serious defect. Do not use device. Call ser-
vice!
Valve error! Do not use device. Call ser-
vice!
Do not use device.
Call service!
Calibration error! The device must be re-
calibrated. Call service!
The device is not calibrated properly.
The device must be re-calibrated. Call ser-
vice!
Contamination! Fluid has penetrated
the Control Unit internally! Ongoing sur-
gery can be finished! Do not use device
for next procedure! Contact technical
service!
The device is contaminated with fluid.
The device must be checked by an autho-
rized service technician or clearly marked
with a label referring to the contamina-
tion and then twice enclosed in a safety
foil, sealed, and returned to the manufac-
turer for repairs.
Critical message
Cause
Remedy
Check connection of Tri-Lumen Filtered
Tube Set! Insufflation will shut down in
15s!
Tube set improperly connected.
Connect the tube set correctly.
Cannula partly occluded! Ensure black
lines on distal cannula are visible inside
cavity!
Tube or cannula has a blockage.
Determine the cause for the occlusion.
Check the positioning of the cannula.
Filter contaminated! Insert new tube
set!
High fluid level of the fluid trap in the
filter housing
Please insert a new tube set.
Fluid Threshold in Filter Exceeded.
AirSeal Insufflation Stopped. Change
Tube Set to Complete Current Proce-
dure. Inspection Required to Assess Pos-
sible Damage to iFS Unit. Call for
Service.
High fluid level of the fluid trap in the
filter housing during AirSeal mode.
Finish procedure and call service.
Fluid Threshold in Filter Exceeded.
Smoke Evacuation Function Stopped.
Change Tube Set to Complete Current
Procedure. Inspection Required to
Assess Possible Damage to iFS Unit. Call
for Service.
High fluid level of the fluid trap in the
filter housing during Smoke Evacua-
tion mode.
Finish procedure and call service.
Change gas bottle! Insufflation stopped!
Low insufflation gas supply (bottle or
house supply) during insufflation.
Check gas supply.
Change gas bottle!
Low insufflation gas supply (bottle or
house supply).
Check gas supply.
Change gas bottle! AirSeal Insufflation
shut down in 100 s! Pressure limited to
12 mmHg!
Low insufflation gas supply (bottle or
house supply) during AirSeal® insuffla-
tion.
Check gas supply.
Insufficient gas! Check gas supply and
restart device!
Device has no gas supply.
Check gas supply.
Check gas supply!
Device has no gas supply.
Check gas supply.
Check gas supply! Insufflation stopped!
Low insufflation gas supply (bottle or
house supply) during insufflation.
Check gas supply.
Check gas supply! AirSeal Insufflation
will stop in 100 s! Pressure limited to
12 mmHg!
Low insufflation gas supply (bottle or
house supply) during AirSeal® insuffla-
tion.
Check gas supply.
Gas level low. Prepare to change gas
bottle.
Low insufflation gas supply (bottle or
house supply) during insufflation.
Check gas supply.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...