Using and Controlling the AirSeal® iFS in its Different Modes
29
EN
ble during insufflation or while insufflation is stopped.
Setting nominal flow:
The device features three different flow level rates that can be adjusted in the
user menu (see chapter 8 "User Menu" for additional information). The facto-
ry defaults are:
Level 1 -> 5 l/min
Level 2 -> 20 l/min
Level 3 -> 40 l/min
• To set the flow rate, press the key
(1)
,
(3)
or
(4)
.
Setting the nominal pressure:
Press the key
»
or
º
(
(5)
or
(8)
) to set the nominal pressure. Values may
range from
5 to max. 20 mmHg
and can be adjusted in increments of 1. The
starting value for the pressure can be set in the user menu within a range of
5 - 15 mmHg.
• Scrolling is enabled by keeping the
»
or
º
key depressed longer than 1.5
seconds.
Increasing the nominal pressure to > 15 mmHg:
Status line depicts the message:
Safety threshold
:
Pressure >15 mmHg!
This is where the recommended range for intra-abdominal and colorectal
pressure ends. Release the key for 2 seconds, then pressure can be increased
above 15 mmHg.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
5. Start insufflation with the
START
key.
The status line depicts
Initial Insufflation
.
The initial insufflation ends when
the nominal pressure value has been reached for the first time. The status
line depicts
Standard Insufflation Active.
NOTE!
For the safety of the patient, it is advised to start intra-abdominal insufflation
with a Veress needle and at the lowest flow rate (level 1, key
(1)
). Exchange the
Veress needle with a conventional cannula once the nominal pressure is reached
and the initial insufflation phase ends.
CAUTION!
Always use Single Lumen Filtered Tube Set with filter (see 5.5.2 "Insufflation
Tube Sets").
WARNING!
Positioning the patient
Always position the patient lower than the device to prevent body fluids from
leaking into the insufflation tube. Actual pressure may increase and fluid may
penetrate the insufflation tube if the patient is repositioned during surgery. If
this occurs, immediately disconnect the insufflation tube. When the patient is
repositioned onto his or her side, internal tissue may block the insufflation chan-
nel. Always insufflate through the elevated side of the patient.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...